The DIAsource CT-U.S.-ELISA is a solid phase Enzyme Amplified Sensitivity Immunoassay performed on microtiterplate. Calibrators and samples react with the capture monoclonal antibody (MAb 1) coated on microtiter well and with a monoclonal antibody (MAb 2) labelled with horseradish peroxidase (HRP). After an incubation period allowing the formation of a sandwich: coated MAb 1 – human CT – MAb 2 – HRP, the microtiterplate is washed to remove unbound enzyme labelled antibody. Bound enzyme-labelled antibody is measured through a chromogenic reaction. Chromogenic solution (TMB ready for use) is added and incubated. The reaction is stopped with the addition of Stop Solution and the microtiterplate is then read at the appropriate wavelength. The amount of substrate turnover is determined colourimetrically by measuring the absorbance, which is proportional to the Calcitonin concentration.A calibration curve is plotted and Calcitonin concentration in samples is determined by interpolation from the calibration curve.
Catalog # | KAR0421 |
Format | ELISA |
Label | HRP |
Size | 96 tests |
Sample Type | Serum |
Sample Volume | 100 µL |
Controls | 2 levels |
Range | 10-400 pg/mL |
Sensitivity | 0,7 pg/mL |
Incubation | Overnight at 2-8°C |
Shelf Life (weeks) | 48 |
Calibration | 1 pg = 0.19µIU NIBSC 89/620 |
For Japan and Brazil, specific registration requirements are necessary. For more information, please contact: regulatory.affairs@diasource.be.